| TestEquity F4 Programmable Controller |
The controller is packaged with a NEMA 4X front face to withstand harsh environments. It is manufactured for TestEquity by Watlow and backed by a three-year warranty. Set Point Ramp
and Soak Programming A profile guide leads the programmer through the programming process by offering choices for step configuration. For better operator recognition profiles and digital I/O used for events can be given names of up to 10 characters in length. Five step types including ramp, soak, jump, auto start and end offer the programmer complete flexibility. Five step types including ramp, soak, jump, auto start and end offer you complete flexibility. Ramp steps can be based on time or rate. Ramp and soak steps can be programmed to wait for up to four event inputs and three process variables. Up to seven event outputs are step selectable. Five sets of PID heat/cool parameters are also step selectable. The auto start step can start a profile based on a set date, a day of the week, or daily. The jump step can be used to jump within a profile or to another profile. Serial Communication Alarms Optional Analog
Retransmit Optional GPIB
Interface Optional Ethernet
Interface |
Temperature Controller Specifications |
|
| Accuracy & Sensor Conformity* | ±0.55°C,
±1 LSD (above -50°C) ±0.65°C, ±1 LSD (below -50°C) |
| Stability | ±0.1°C/°C rise in ambient |
| Digital Inputs | Four inputs. Contact closure or dc voltage, 10 kohm impedance |
| Retransmit
Outputs (Optional) |
Two outputs. User-selectable ranges: 0 to 10 VDC, 0 to 5 VDC, 1 to 5 VDC 0 to 20 mA, 4 to 20 mA |
| Alarm Outputs | Two alarm
outputs. Electromechanical relay, Form C, 2 A @ 20 VDC or 240 VAC max. |
| Digital Outputs | Seven
open-collector ouputs available for customer use. OFF: 42 VDC @ 10 µA max. ON: 0.2 VDC @ 50 mA sink max. Internal supply: 5 VDC @ 80 mA |
| Communications | RS-232 and RS-485 serial communications with Modbus RTU protocol |
| Safety & Agency Approvals | UL/c-UL
916-listed, File #E185611 CE to EN61010 NEMA 4X and IP65 CE EMC to EN50082-2 CE EMC to EN55011 |
| Displays | Process:
5, seven-segment red LED Interface Display: 4-line high-definition green LCD Selectable °C or °F |
| Data Retention | Retention upon power failure via nonvolatile memory (seven years for battery-backed RAM) |
| *Note: Total system accuracy in the chamber includes thermocouple wire accuracy. Thermocouple wire accuracy is ±1°C or 0.75% of reading, whichever is greater. | |